SmartPro 120V 700VA 450W Medical-Grade Line-
Interactive Tower UPS with 4 Outlets, Full
Isolation, USB, DB9
MODEL NUMBER: SMART700HG
Provides UL 60601-1 compliant, medical-grade 700VA/450W power backup, ideal for use in patient-care
vicinities. The built-in isolation transformer offers full isolation while maintaining AC surge protection in all
modes.
Description
Tripp Lite’s SMART700HG medical-grade UPS system provides dependable battery backup and is fully
compliant with UL 60601-1 regulations when plugged into a hospital-grade outlet, making it ideal for use in
patient-care vicinities. The 700VA/450W capacity provides up to 42 minutes of runtime for a half load, or
18 minutes for a full load so you can work through short power failures and perform an orderly system
shutdown in a prolonged blackout.
Built-in line-interactive voltage regulation corrects brownouts as low as 81V and overvoltages as high as
143V to normal 120V nominal output, without drawing battery power. The SMART700HG’s built-in
isolation transformer with Faraday shield offers full isolation and removes common mode noise while
maintaining full AC surge suppression in all modes.
This UPS system features four NEMA 5-15R-HG hospital-grade outlets and has a hospital-grade NEMA
5-15P-HG plug with a 6 ft. cord for flexibility in placement. All four outlets provide both backup protection
and surge protection, rated to 480 joules.
The SMART700HG also features diagnostic LEDs, as well as a USB port and a DB9 serial port that work
with Tripp Lite’s free downloadable PowerAlert software for safe, unattended system shutdown and file
saves in the event of a prolonged power failure.
Application Usage
This UPS contains a lead acid battery designed to be connected continuously to an AC power source,
except during brief power outages. It is not intended for frequent charge/discharge cycles, such as mobile
cart applications where the UPS is often disconnected from the power source. Using the SMART700HG in
this type of application will void the battery warranty. For mobile applications, Tripp Lite recommends the
SMART700HGL with lithium iron phosphate (LiFePo4) battery.
Features
Provides Ideal UPS Backup in Healthcare Facilities
Highlights
Supports a half load up to 42
minutes; full load up to 18
minutes
Four hospital-grade outlets and
hospital-grade plug
480 joule surge suppression
rating protects connected
equipment
Alarm mute/alarm quiet control
is ideal for use around patients
Approved for use within the
Patient Care Vicinity
Approved to EN60601-1-2:2015
& IEC60601-1-2:2014, 4th
Edition
Package Includes
SMART700HG 700VA/450W
Medical-Grade Tower UPS
System
USB monitoring cable, 6 ft.
DB9 serial monitoring cable, 6 ft.
Owner’s Manual
1 / 5
Specifications
Compliant with UL 60601-1 standards for use in patient-care vicinities, when plugged into a hospital-
grade outlet
Built-in isolation transformer with Faraday shield offers full isolation and removes common mode noise
Hospital-grade plug with 6 ft. cord provides flexibility in placement so healthcare workers can focus on
treating the patient
Battery Backup and Full Surge Protection
42 minutes runtime for half load (225W); 18 minutes for full load (450W)
All four hospital-grade outlets provide surge protection rated to 480 joules
EMI/RFI noise filtering ensures equipment runs smoothly without line noise interference
Resistant to electrostatic discharge
LEDs and Audible Alarms Keep You Aware of Power Status
Indicator lights on the front of the unit show you when the UPS is on and supplying utility power, when
operating in voltage correction mode or in battery mode, relative load level and low-battery warning
Audible alarms alert you to power failure, overload and low battery conditions; the alarm’s mute/quiet
mode ensures patients are not disturbed
Saves Your Files, Even When You’re Away
Works with Tripp Lite PowerAlert software (available as a free download) to enable safe, unattended
system shutdown and files saves during prolonged power failures
Connects to your computer via either the DB9 port or USB port
OVERVIEW
UPC Code 037332095107
UPS Type Line-Interactive
OUTPUT
Output Volt Amp Capacity (VA) 700
Output Watt Capacity (Watts) 450
Power Factor 0.6
Frequency Compatibility 60 Hz
Output Voltage Regulation (Line
Mode) -18%, +8%
Output Voltage Regulation (Battery
Mode) 115V (+/- 5%)
Output Receptacle Details Hospital grade output receptacles
2 / 5
TRIPP°LITE
Output AC Waveform (AC Mode) Sine wave
Output AC Waveform (Battery
Mode) PWM sine wave
Nominal Output Voltage(s)
Supported 115V; 120V
Output Receptacles (4) 5-15R
Individually Controllable Load Banks No
INPUT
Rated input current (Maximum Load) 5.5A
Nominal Input Voltage(s) Supported 120V AC
UPS Input Connection Type 5-15P-HG
Input Circuit Breakers 10A
UPS Input Cord Length (ft.) 6
UPS Input Cord Length (m) 1.8
Recommended Electrical Service 15A 120V
Input Phase Single-Phase
Input Cord Length (ft.) 6
Input Cord Length (m) 1.83
BATTERY
Full Load Runtime (min.) 18 min. (450w)
Half Load Runtime (min.) 42 min. (225w)
DC System Voltage (VDC) 36
Battery Recharge Rate (Included
Batteries) Less than 4 hours from 10% to 90%
Internal UPS Replacement Battery
Cartridge RBC36SLT
Battery Access Battery access door
Battery Replacement Description Hot-swappable, user replaceable batteries
Expandable Runtime No
VOLTAGE REGULATION
Voltage Regulation Description Automatic voltage regulation (AVR) maintains line power operation with an input voltage range of 81 to 143
Overvoltage Correction Input voltages between 126 and 143 are reduced by 9%
Undervoltage Correction Input voltages between 96 and 106 are boosted by 10%
Severe Undervoltage Correction Input voltages between 81 and 95 are boosted by 21%
USER INTERFACE, ALERTS & CONTROLS
3 / 5
TRIPP°LITE
Switches 2 Switches control off/on power status and alarm-cancel/self-test operation
Alarm Cancel Operation Power-fail alarm can be silenced using alarm-cancel switch; once silenced, alarm will re-sound to indicate low-
battery status
Audible Alarm Audible alarm indicates power-failure, overload and low-battery conditions
LED Indicators 5 LEDs indicate line power, battery power, overload, voltage regulation and battery low/replace status
SURGE / NOISE SUPPRESSION
UPS AC Suppression Joule Rating 480
UPS AC Suppression Response
Time Instantaneous
EMI / RFI AC Noise Suppression Yes
AC Suppression Joule Rating 480
AC Suppression Response Time Instantaneous
ISOLATION
Output Isolation Transformer Hospital grade load isolation transformer
Leakage Current Equal to 100 microamps
PHYSICAL
Cooling Method Fan
Installation Form Factors Supported
with Optional Accessories Wallmount (UPSWM - tower mounting kit)
Primary Form Factor Tower
Primary UPS Depth (mm) 229
Primary UPS Height (mm) 325
Primary UPS Width (mm) 191
Shipping Dimensions (hwd / cm) 43.18 x 28.45 x 33.02
Shipping Dimensions (hwd / in.) 17.00 x 11.20 x 13.00
Shipping Weight (kg) 20.82
Shipping Weight (lbs.) 45.90
UPS Housing Material PVC
UPS Power Module Dimensions
(hwd, cm) 32.51 x 19.05 x 22.86
UPS Power Module Dimensions
(hwd, in.) 12.8 x 7.5 x 9
UPS Power Module Weight (kg) 20.05
UPS Power Module Weight (lbs.) 44.2
ENVIRONMENTAL
4 / 5
TRIPP°LITE
© 2019 Tripp Lite. All rights reserved. All product and company names are trademarks or registered trademarks of their respective holders. Use of them does
not imply any affiliation with or endorsement by them. Tripp Lite has a policy of continuous improvement. Specifications are subject to change without notice.
Tripp Lite uses primary and third-party agencies to test its products for compliance with standards. See a list of Tripp Lite's testing agencies:
https://www.tripplite.com/products/product-certification-agencies
Operating Temperature Range +32 to +104 degrees Fahrenheit / 0 to +40 degrees Celsius
Storage Temperature Range +5 to +122 degrees Fahrenheit / -15 to +50 degrees Celsius
Relative Humidity 0 to 95%, non-condensing
AC Mode BTU / Hr. (Full Load) 97.7
Battery Mode BTU / Hr. (Full Load) 189.1
AC Mode Efficiency Rating (100%
Load) 94%
COMMUNICATIONS
Communications Interface USB; DB9 Serial
Network Monitoring Port Description Supports detailed monitoring of UPS and site power conditions
PowerAlert Software For local monitoring via the UPS’s built-in communication ports , download PowerAlert Local software at
http://www.tripplite.com/poweralert
Communications Cable USB and DB9 serial cabling included
LINE / BATTERY TRANSFER
Transfer Time 2-4 milliseconds
Low Voltage Transfer to Battery
Power (Setpoint) 81
High Voltage Transfer to Battery
Power (Setpoint) 143
SPECIAL FEATURES
Grounding Lug Grounding lug
Cold Start (Startup in Battery Mode
During a Power Failure) Cold-start operation supported
High Availability UPS Features Hot swappable batteries
Green Energy-Saving Features Greater than 95% efficiency - GREEN UPS
CERTIFICATIONS
UPS Certifications Meets FCC Part 15 Category A (EMI); ROHS (Restriction of Hazardous Substances); Tested to CSA (Canada);
Tested to UL1778 (USA); Tested to UL60601-1 (medical grade)
WARRANTY
Product Warranty Period
(Worldwide) 2-year limited warranty